ClearanceBrief
Recall Z-1037-2013 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

SIEMENS brand COHERENCE RT Therapist versions RTT MR1.5, RTT 2.1/2.1A and RTT 2.2/ 2.2B; Software release for COHERENCE Therapist 2.3 for Primus and Oncor Linear Accelerators; Model: COHERENCE TH, Part No. 7339125, - Current R630; Product Usage: The intended use of the SIEMENS branded ONCOR and PRIMUS family of linear accelerator systems is to deliver X-ray photon and electron radiation for

DateApr 2, 2013 (initiated Mar 5, 13)
StatusTerminated
ReasonSiemen's issued a Customer Information Letter under Update Instruction TH016/12S to alert all its affected customers of the software issues. The software release is a result of the final system testing addressing multiple performance defects and several safety issues that were not considered critical (for immediate correction) for the safe use of the device.
Root causeSoftware design (manufacturing process)
ActionSiemens Medical Solutions sent a Customer Information letter via mail to all affected customers. The letter identified the affected product, problem and provide instructions on the correction. Customers were instructed to include the Customer Information letter in the Owner Manual chapter "Safety Advisory Letters". 9-16-13 UPDATE: - Control Console SW 9.2.18 was released to support the rol-out 10/18/2013; Distribution of update instructions for TH017/13S and TH018/13S - Distribtuion of the Control Console SW 9.2.18 released
Quantity621 units (foreign account) in total
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060226ONCOR EXPRESSION
Siemens Medical Solutions USA, Inc., Mar 15, 2006
K092145160MLC OPTION
Siemens Medical Solutions USA, Inc., Sep 18, 2009
K123812COHERENCE RT THERAPIST CONNECT WORKSPACE, V2.3, ARIA CONNECTIVITY OPTION FOR SIEMENS NON-IMAGING SYSTEMS, ARIA CONNECTIV
Siemens Medical Solutions USA, Inc., Mar 1, 2013
K993425PRIMUS
Siemens Medical Solutions USA, Inc., Nov 26, 1999