Recall Z-0994-2013 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Toshiba American Medical Systems Inc

Toshiba Infinix-i X-Ray System. A radiographic and fluoroscopic x-ray system for use in diagnostic and angiographic procedures.

DateMar 25, 2013 (initiated Nov 8, 2012)
StatusTerminated
ReasonLost and incorrect images. When using the "F-Rec" fluoroscopy image acquisition recording option in the "manual mode" an infrequent software error may cause images from another study to be displayed on the monitor. The acquired images may be lost requiring additional image acquisitions to be performed.
Root causeSoftware design
ActionIn November, 2012, Toshiba America Medical Systems, Inc. sent an Urgent: Medical Device Correction letter to customers. The letter contains a return reply form that is to be faxed to TAMS for retention.
Quantity95 units
DistributionUSA nationwide. No international.
FDA recordOpen on fda.gov