Recall Z-1068-2013 ยท JAK System, X-Ray, Tomography, Computed

Toshiba American Medical Systems Inc

Toshiba TSX-30IC AQUlLlON ONE System, Computed Tomography X-Ray. This system supports Whole body scanning.

DateApr 8, 2013 (initiated Nov 14, 2012)
StatusTerminated
ReasonDue to a problem with the control software of the x-ray high-voltage generator, if scanning is performed with a tube current less than 30mA, the control software may detect an error and perform error processing even though the x-ray output is normal.
Root causeProcess design
ActionToshiba American Medical Systems Inc. sent a Urgent Medical Device Correction letter dated November 14, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customer was informed a New software to prevent the occurrence of this issue will be installed on your system. Toshiba Service Representative will contact you for an appointment to install the New Software on your Aquilion One System. Please fax the return reply form to (877) 349-3054. We thank you for your urgent attention to this matter. If you have any fu
Quantity1 unit
DistributionDistributed in the state of MD.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122109AQUILION ONE VISION, V4.90
Toshiba America Medical Systems, In.C, Sep 21, 2012