Recall Z-0997-2013 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray system

DateMar 26, 2013 (initiated Mar 8, 2013)
StatusTerminated
ReasonThe firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may lead to imprecise registration of the DSA mask image with the roadmap image displayed on the live monitor.
Root causeSoftware design
ActionSiemens sent a Customer Safety Advisory Notice on March 8, 2013, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to pass the information on to all those who need to be aware of it and to instruct their personnel accordingly. Customers were asked to forward the safety information to other organizations that couald be affected by this action. For questions regarding this recall call 610-219-6300.
Quantity81
DistributionNationwide Distribution including AL, CA, CO, FL, IA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, ND, NJ, NY, OH, PA, SC, TX, UT, VA, WA, and WI and to Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073290ARTIS ZEE AND ZEEGO
Siemens Medical Solutions USA, Inc., Feb 11, 2008
K090745ARTIS ZEE / ZEEGOS WITH SW VC14
Siemens Medical Solutions USA, Inc., Jun 18, 2009