Recall Z-1074-2013 ยท NBW System, Test, Blood Glucose, Over The Counter

Lifescan Inc

All OneTouch Verio IQ Blood Glucose Meters sold as: Verio IQ System Kits - Meter; Verio IQ Starter kits - Meter; Verio IQ Warranty Meter. Product Usage: The OneTouch VerioIQ Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The system is intended to be used by a single pati

DateApr 12, 2013 (initiated Mar 11, 2013)
StatusTerminated
ReasonThe Verio IQ meter will shut off and revert to set up mode at glucose values above 1023 instead of displaying EXTREME HIGH GLUCOSE.
Root causeSoftware design
ActionLifescan sent an Urgent Medical Device Voluntary Recall letter dated March 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to identify and hold all affected product in inventory and discontinue distributing. The letter states that a Lifescan sales representative will contact customers to arrange for pick up and replacement.
QuantityUS 710,413
DistributionUSA Nationwide Distribution
FDA recordOpen on fda.gov