Recall Z-0998-2013 ยท MLR Sterilizer, Chemical

Advanced Sterilization Products

STERRAD 200 System Carriage, P/N 10208. Used as a tool to expedite loading and unloading processed items from a STERRAD 200 sterilizer.

DateMar 26, 2013 (initiated Mar 13, 2012)
StatusTerminated
ReasonAdvanced Sterilization Products (ASP) is recalling the STERRAD 200 System Carriage because it can potentially become dislodged when an excessive load is applied.
Root causeDevice Design
ActionAdvanced Sterilization Products (ASP) sent an "URGENT MEDICAL DEVICE NOTIFICATION" letter dated March 13, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The letter informed the customers that ASP recently determined that the push bar located on the STERRAD 200 System Carriage can potentially become dislodged when an excessive load is applied. Customers with questions are instructed to contact Stericycle at (877) 931-0642.
Quantity2,457 units
DistributionWorldwide Distribution-USA (nationwide) and the countries of Egypt, Germany, Italy, France, Saudi Arabia, UAE, Japan, Singapore, China, Thailand, Hong Kong, Hungary, Israel, Slovania, Poland, Czech Republic, Taiwan, Sweden, Greece, Korea, Ireland, and United Kingdom.