ClearanceBrief
Recall Z-1001-2013 ยท OJU Skin Prep Tray

CareFusion 213, LLC

Sepp, Compound Benzoin Tincture Packaged and Tincture 3,000 Bulk. Product Usage: Preoperative skin product

DateMar 26, 2013 (initiated Mar 4, 2013)
StatusTerminated
ReasonAn internal review of CareFusion's labeling for preoperative skin prep product codes was recently conducted. The labeling of the product codes identified above state that the contents of the package are sterile and does not clarify that this sterility claim is only applicable to the applicator. The antiseptic solution that is applied to the skin is a cutaneous disinfectant, and is not sterilized.
Root causeLabeling False and Misleading
ActionCareFusion sent an Urgent Recall Notice dated March 4, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately check inventory, quarantine and return the affected product for replacement. For questions call (913) 345-3504.
Quantity13,749,000
DistributionUSA Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.