Recall Z-1114-2013 ยท GJS Test, Time, Prothrombin

Roche Diagnostics Operations, Inc.

CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").

DateApr 15, 2013 (initiated Mar 6, 2013)
StatusTerminated
ReasonRoche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.
Root causeDevice Design
ActionConsignees of both devices were notified by Urgent Medical Device Correction letters sent via UPS Ground (receipt required) beginning 3/06/2013. The notice contained recommendation not to use EAN-13 barcode system and ensure barcodes used meet quality standards; use operator and patient lists (downloaded from a data manager) in order to check any decoded identification against the respective list. The notice directed forwarding of the notice if the product was further distributed. Customers were asked to complete the attached fax form. Contact the Customer Care Service Center, 24 hours a da
Quantity4467 both devices
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093460COAGUCHEK XS PRO SYSTEM
Roche Diagnostics, Mar 18, 2010