Recall Z-2677-2017 ยท FRN Pump, Infusion

ICU

The Plum A+ Infusion Systems are a multipurpose family of devices that provides precise delivery of multiple therapies across the general spectrum of clinical care. The Plum A+ family infusion pumps are self contained, microprocessor based infusion devices. The pumps are available in single channel and in triple channel pumping mechanism types.

DateJul 7, 2017 (initiated Feb 1, 2013)
StatusTerminated
ReasonA possible interruption of therapy when an E321 error code occurs while using the infuser, which is caused by battery issues.
Root causeOther
ActionHospira has contracted with Stericycle Inc., in Indianapolis, IN to send letters and a reply form to Hospira customers. Receipt of the letters will be confirmed by UPS tracking. Customers will be requested to return a reply form to Stericycle indicating they have received the letter, and notified their customers. The customers will also be asked to supply the number of Plum A+ infusers at their facility.
Quantity71 units
DistributionNationwide, Worldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070398HOSPIRA PLUM A+ INFUSION PUMP, MODEL 12391
Hospira, Inc., Apr 24, 2007