Recall Z-1059-2013 ยท KHE Reagent, Occult Blood

Polymedco, Inc

Test Strips packaged in the OC-Light Manual iFOBT Kit, Catalog No. FOB50. For the rapid, qualitative detection of human hemoglobin in feces.

DateApr 4, 2013 (initiated Feb 12, 2013)
StatusTerminated
ReasonThe test strips in the affected product lots of the OC Light immunological fecal occult blood test are being recalled because they may not be stable through their labeled shelf-life. There may be a possibility of obtaining false negative results when using the affected lots.
Root causeProcess control
ActionPolymedco initiated phone calls to clients on 2/12/13. When the contact was made with the client, an URGENT DEVICE RECALL letter and Fax Form were sent by e-mail/fax. Mass mailing of the URGENT DEVICE RECALL letter dated February 12, 2013 and Fax Form was done by First Class USPS on 2/14/13-2/18/13. The recall affects the test strips only. For questions, please contact Polymedco Technical Support at 1-800-431-2123 ext 285.
Quantity355,350 test strips
DistributionWorldwide Distribution-USA (nationwide) including Puerto Rico, and the countries of Canada and Guam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041297POLYMEDCO OC LIGHT FOBT TEST
Polymedco, Inc., Aug 12, 2004