Recall Z-0968-2013 ยท CBK Ventilator, Continuous, Facility Use

Del Mar Reynolds Medical, Ltd.

Spacelabs BleaseSirius Anesthesia Workstation, BleaseFocus Anesthesia Workstation, and CAS I/II Absorber contained in Service Kits Part Number 12200902, 050-0659-00 and 050-0901-00 and any workstation installed with the CAS I/II Absorber. Anesthesia workstation for use in the hospital environment and locations not requiring portability.

DateMar 20, 2013 (initiated Jan 15, 2013)
StatusTerminated
ReasonA defect in CAS I/II Absorbers in the Spacelabs Anesthesia Workstations and Service Kits may cause a condition leading to an increase in the carbon dioxide concentration within the inspired gas being delivered to the patient.
Root causeProcess control
ActionSpacelabs sent an URGENT-MEDICAL DEVICE CORRECTION letter, dated 16 January 2013, to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers until a Spacelabs representative replaces the affected product: 1) Immediately cease the use of BleaseSirius and BleaseFocus Anesthesia Workstations that shipped after 31 October 2012. 2) Immediately cease use of Service Kits Part Number 12200902, 050-0659-00 and 050-0901-00 and any workstation containing a CAS I/II Absorber from such Service Kit. 3) Advise your staff
Quantity108 total (17 units in the US and 91 units outside the US)
DistributionWorldwide Distribution - USA Nationwide including the states of New Mexico, North Carolina and the countries of Argentina, China, Mexico, and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003251BLEASE FRONTLINE PLUS RANGE, ANESTHESIA MACHINES, MODEL 440, 560 AND 690
Blease Medical Equipment , Ltd., May 23, 2001
K101850SPACELABS BLEASESIRIUS ANESTHESIA WORKSTATION
Spacelabs Healthcare, Ltd., Mar 2, 2011