Recall Z-1113-2013 ยท NBW System, Test, Blood Glucose, Over The Counter

Roche Diagnostics Operations, Inc.

ACCU-CHEK Inform II Blood Glucose Monitoring System bar code scanner. Used with the ACCU-CHEK Inform II System to quantitatively measure glucose (sugar).

DateApr 15, 2013 (initiated Mar 6, 2013)
StatusTerminated
ReasonRoche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards.
Root causeDevice Design
ActionConsignees of both devices were notified by Urgent Medical Device Correction letters sent via UPS Ground (receipt required) beginning 3/06/2013. The notice contained recommendation not to use EAN-13 barcode system and ensure barcodes used meet quality standards; use operator and patient lists (downloaded from a data manager) in order to check any decoded identification against the respective list. The notice directed forwarding of the notice if the product was further distributed. Customers were asked to complete the attached fax form. Contact the Customer Care Service Center, 24 hours a da
QuantityBoth devices 4467
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K121679ACCU-CHEK INFORM II BLOOD GLUCOSE MONITORING SYSTEM
Roche Diagnostics, Oct 11, 2012