Recall Z-1064-2013 ยท DSP System, Balloon, Intra-Aortic And Control

Arrow International Inc

Arrow International FiberOptix Intra-Aortic Balloon Catheter Kit (IAB) 8 Fr 30 cc; Product Number: IAB-05830-LWS. The IAB is utilized for intra-aortic balloon counter-pulsation therapy.

DateApr 6, 2013 (initiated Nov 27, 2012)
StatusTerminated
ReasonPotential to have an internal leak at the fill port of the stainless steel tip, which may necessitate removal and replacement of the catheter.
Root causeProcess control
ActionArrow International issued Urgent Medical Device Recall notification on 11/27/12. The letter identified the affected product and the reason for the recall. Accounts are requested to check their stock for the products included within the scope of this recall. Also, customers should cease use and distribution, and quarantine all affected product immediately. Contact Arrow's Customer Service Department at 800-523-8446 for a Return Authorization Number. Customers are to complete and fax the Recall Acknowledgement & Stock Status Form to the number provided, and return the affected product. If
Quantity188 units
DistributionWorldwide Distribution -- USA, Belgium, Great Britain, Italy, Kazakhstan, Latvia, Russia, Slovenia, Turkey, and South Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K000729ARROW ULTRA 8 INTRA-AORTIC BALLOON CATHETERS (IABS) 8FR 30CC AND 40CC UNIVERSAL
Arrow Intl., Inc., May 19, 2000
K021462ARROW INTRA-AORTIC BALLOON (IAB) CATHETER WITH A FIBER OPTIC SENSOR AND A MEASUREMENT SYSTEM, MODELS IAB-05830-F,IAB5840
Arrow Intl., Inc., Jun 6, 2002