Recall Z-0955-2013 ยท DTL Adaptor, Stopcock, Manifold, Fitting, Cardiopulmonary Bypass

Vascular Solutions, Inc.

Vascular Solutions, Inc., Guardian II and Guardian II NC Hemostasis Valve, used in cardiopulmonary bypass. Model numbers: FH101, FH101-T, FH101-25, FH101-50, 8210, 8211, FH102, FH102-T, FH102-25, FH102-50, 8215, 8216, 8241, 8242, 8244, 8272 and 8274. Vascular Solutions Zerusa Limited. Intended to maintain hemostasis during the use of diagnostic/interventional devices during diagnostic/interve

DateMar 20, 2013 (initiated Feb 27, 2013)
StatusTerminated
ReasonRisk that air may be introduced into the device which may lead to an air embolism.
Root causeVendor change control
ActionConsignees were sent a Vascular Solutions "Urgent Medical Device Recall" letter dated February 28, 2013. The letter described the problem and the product involved in the recall. In addition, consignees were advised to identify the location of and immediately remove the recalled product from their inventory and secure it. Consignees were requested to complete and return the VSI Account Inventory Form. The letter also stated that VSI Customer Service would contact customers within 1-2 business days after receiving the completed VSI Account Inventory Form. All returned affected product will
Quantity79,869
DistributionWorldwide Distribution -- USA, including the states of AL, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, IL IN, IA, KS, KY, LA, MD, MA, MI, MN, MS, MO, NE, NV, NJ, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT , VT, VA, WA, WV, and WI, and countries of GREECE, DENMARK, AUSTRIA, UNITED KINGDOM, SWITZERLAN
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092711GUARDIAN II HEMOSTASIS VALVE
Zerusa Limited, Oct 1, 2009
K101113GUARDIAN II NC HEMOSTASIS VALVE
Zerusa Limited, May 21, 2010
K122301GUARDIAN II HEMOSTASIS VALVE GUARDIAN II NC HEMOSTASIS VALVE
Vascular Solutions Zerusa , Ltd., Sep 7, 2012