Recall Z-1183-2013 ยท FKQ System, Dialysate Delivery, Central Multiple Patient

B. Braun Medical, Inc.

Diacap(R) Ultra Dialysis Fluid Filter, Catalog Number 7107367. Intended to filter bacteria and endotoxins from dialysis fluid.

DateApr 26, 2013 (initiated Mar 1, 2013)
StatusTerminated
ReasonThere is the potential of the residual moisture of the Diacap Ultra membrane being reduced over time, which results in the reduction of the membrane permeability (ultrafiltration coefficient).
Root causeProcess design
ActionB. Braun Medical Inc. sent and "URGENT RECALL NOTICE" dated February 28, 2013 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the Customer Support Department at (800) 848-2066 for questions regarding this issue. This recall was expanded on/about June 18, 2013 to include additional lot #s 305010113 and 305000113. Customers were notified via certified mail on/about June 14, 2013.
Quantity1580
DistributionDistribution and manufacturing dates for the expansion lots was reported as 5/23-6/14/2013 and 1/22-24/2013 respectively. Nationwide Distribution including the states of AZ, CA, DE, FL, GA, HI, IA, ID, IL, KS, KY, LA, MN, MO, NJ, NM, NY, OH, OR, PA, SC, SD, TN, TX, UT, WA and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052764DIACAP ULTRA DIALYSIS FLUID FILTER