Recall Z-1031-2013 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Medtronic Neuromodulation

Medtronic Prostiva RF Therapy Generator. The Prostiva RF system is indicated for the treatment of symptoms due to urinary outflow obstruction secondary to benign prostatic hyperplasia (BPH) in men over the age of 50 with prostate sizes between 20 and 50 cm3.

DateApr 2, 2013 (initiated Feb 27, 2013)
StatusTerminated
ReasonMedtronic has found through device testing that if the optional footswitch is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the footswitch connector port may toggle the RF output on or off. If this were to occur, it could result in unintentionally starting or stopping the lesion process, depending on the current status of the equi
Root causeNonconforming Material/Component
ActionMedtronic sent a "Urgent Medical Device Correction" letter dated January 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Medtronic has found through device testing that if the optional footswitch (Model 60883) is not connected to the Model 8930 RF generator during use, an electrostatic discharge (i.e. static electricity) to the footswitch connector port may toggle the RF output on or off. If this were to occur, it could result in unintentionally starting or stopping the lesion process, depending on the curren
Quantity848 (696 US, 152 OUS)
DistributionNationwide Distribution including Puerto Rico and all States in continental USA except ID, MT, and NH.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K113380PROSTIVA RF THERAPY HAND PIECE PROSTIVA RF THERAPY RF GENERATOR
Medtronic, Inc., Nov 15, 2012