Recall Z-1024-2013 ยท OPP Pump, Infusion, Insulin Bolus

Zyno Medical LLC

Z-800 Volumetric Infusion System. For intravenous infusion of parenteral fluids, blood and blood products to a human patient.

DateMar 29, 2013 (initiated Jun 13, 2011)
StatusTerminated
ReasonUpdate Instruction For Use for Preventative Maintenance (PM) Schedule and addition of Odometer software to monitor accumulative fluid volume infused by the pump
Root causeProcess design
ActionZyno Medical initiated recall 6/13/11 to accounts advising of a Preventative Maintenance (PM) Schedule and a software upgrade Odometer to monitor counts of accumulative fluid volume infused by the unit. On 3/1/13, firm issued correction notice to users that have not had the software implemented on their unit. Units will be serviced at their scheduled PM. The notice identified the product, problem, and actions to be taken by the customers.
Quantity2589 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071545Z-800 VOLUMETRIC INFUSION PUMP
Zyno Medical, LLC, Oct 31, 2007
K100705ZYNO Z-800 INFUSION SYSTEM MODEL Z-800
Zyno Medical, LLC, Jul 8, 2010