Recall Z-0939-2013 ยท MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Draeger Medical, Inc.

Draeger Infinity Acute Care System (IACS) Monitoring Solution. Catalog Number MS20783 (Infinity MCable-Dual Hemo) Physiological patient monitoring system.

DateMar 11, 2013 (initiated Dec 13, 2012)
StatusTerminated
ReasonThe displayed value for the invasive blood pressure measurement of the IACS was different than the actual patient values. Liquid ingress in the cable connection may cause inaccurate measurements. This can happen if the Dual Hemo MCable is used in a non-vertical position, a vertical position but upside down, or laying it on the floor or on the patient bed.
Root causeLabeling design
ActionDrager sent an "URGENT MEDICAL DEVICE RECALL" letter dated December 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 1-800-543-5047 for questions relating to this notice.
Quantity260
DistributionDistributed in the states of CA, IL, and MA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K093788INFINITY ACUTE CARE SYSTEM (IACS) MONITORING SOLUTION
Draeger Medical Systems, Inc., Mar 8, 2010
K113798INFINITY ACUTE CARE SYSTEM, INFINITY MCABLE- MASIMO RAINBOW SET
Draeger Medical Systems, Inc., Mar 22, 2012