Recall Z-0915-2013 ยท FLE Sterilizer, Steam

Steris Corporation

AMSCO C and AMSCO 400 Steam Sterilizers designed for sterilization of heat and moisture-stable materials used in healthcare facilities.

DateMar 6, 2013 (initiated Feb 12, 2013)
StatusTerminated
ReasonThe control boards alarm set points and temperature tolerances allowed for erroneous activation of the AD (analog/digital) alarm as temperature fluctuated during operation. When this alarm occurs, the sterilizer locks up and cannot be used until serviced, causing Customer nuisance.
Root causeSoftware design
ActionSteris Corporation initiated their recall on February 12, 2013. They will notify customers via onsite visits by STERIS Field Service Technicians to each affected Customer site. For further questions, please call (440) 392-7601.
Quantity395 Units