ClearanceBrief
Recall Z-1063-2013 ยท GDW Staple, Implantable

Covidien LP

Covidien Endo GIA AutoSuture Universal Loading Unit, 30mm-2.0mm Product Code: 030450 Applications in abdominal, gynecologic, pediatric anti thoracic surgery for resection, transection and creation of anastomosis.

DateApr 5, 2013 (initiated Feb 11, 2013)
StatusTerminated
ReasonSingle Use Loading Unit contained two staples loaded in each cartridge pocket and may result in difficulty firing and removing the device from the application site, which may require medical intervention
Root causeProcess control
ActionCovidien notified customers with an Urgent Medical Device Recall notification on February 11, 2013. The letter identified the affected product and the reason for the recall. Accounts were requested to quarantine and discontinue use of the affected devices, and notify Universal Loading Unit 30mm-2.0mm users of this recall. Customers should follow the instructions provided in order to return the affected product. Customers are to complete and return the Product Return Form regardless if they have any affected product or not. For questions regarding the RGA I return process, please contact
Quantity6,751 units
DistributionWorldwide Distribution -- USA, Asia Pacific, Canada, Europe, and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K111825ENDO GIA(TM) STAPLERS, DST SERIES(TM) GIA(TM) STAPLERS, DST SERIES(TM) TA(TM) STAPLERS
Covidien, Formerly US Surgical A Divison of Tyco H, Oct 25, 2011