Recall Z-0899-2013 ยท JAK System, X-Ray, Tomography, Computed

Philips Medical Systems (Cleveland) Inc

Philips MX 16-Slice CT Scanner System, a whole-body computed tomography X-ray system.

DateFeb 28, 2013 (initiated Dec 19, 2012)
StatusTerminated
ReasonThe four screws that secure the patient table top to the carrier became loose, which caused the table top to detach.
Root causeComponent design/selection
ActionPhilips Healthcare sent an Urgent - Medical Device Correction letter dated December 14, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to have BioMedical Engineering personnel inspect the patient table top screws that secure it to the carrier. Customers were instructed that If it is found that the table top screws are improperly seated, they should stop using the system and contact their local Philips Field Service Engineer for repair. A Philips Field Service Engineer will schedule a tim
Quantity412 Units
DistributionWorldwide Distribution - USA including AL, AR, CA, FL, GA, IL, IN, KY, MI, MS, NC, NJ, NV, NY, OH, PA, PR and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083498PHILIPS MODEL MX 16 SLICE CT SYSTEM
Philips Medical Systems (Cleveland), Inc., Dec 17, 2008