Recall Z-1050-2013 ยท JQP Calculator/Data Processing Module, For Clinical Use

Horiba Instruments, Inc dba Horiba Medical

ABX Micros IM2, Part# M60CSB110EN04 and D00A00202 The IM2 Data Management System consists of the IM2 Data Manager Software, Desktop Computer, keyboard, mouse, and a unidirectional serial cable. The IM2 software is used to store, retrieve, and process laboratory data received via a unidirectional serial connection from the ABX Micros 60.

DateApr 3, 2013 (initiated Feb 15, 2013)
StatusTerminated
ReasonHoriba Medical is recalling ABX Micros IM2 Data Management System because it may display incorrect patient test results under unanticipated operating conditions.
Root causeDevice Design
ActionHoriba Medical sent an Urgent Safety Notice dated February 15, 2013, to all affected customers. The letter informed the customers that Horiba Medical is informing all ABX Micros IM2 Data Management System customers of their intention to remove the ABX Micros IM2 product from the market. The letter informed the customers of the problem identified and the action to be taken. Customers were instructed to complete and return the enclosed response form within 10 days. For questions regarding this recall call 949-453-0500, ext 208.
Quantity180 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall