Recall Z-0948-2013 ยท JAK System, X-Ray, Tomography, Computed

Siemens Medical Solutions USA, Inc

Siemens Somatom CTs equipped with software syngo CT2010A, syngo CT2010B, syngo CT2011A and CT2012B. Computed tomography x-ray system.

DateMar 14, 2013 (initiated Nov 27, 2012)
StatusTerminated
ReasonSporadic hang-up malfunctions concerning 3D reconstructions during activated "Preview Image" functionality.
Root causeSoftware design
ActionSiemens sent a "CUSTOMER SAFETY ADVISORY NOTICE" dated November 27, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. The correction for this problem will be available with Service Pack 1 for syngo CT2011A And Service Pack 4 for syngo CT2012B. Both Service pack are planned to be released starting February 2013. Systems still running on syngo CT2010A or syngo CT2010B are already scheduled to be upgraded to syngo CT2011A or syngo CT2012B respectively. Contact your local Siemens Application Specialist for additional informati
Quantity896
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K122471SOMATOM DEFINITION
Siemens Medical Systems, Inc., Sep 11, 2012