Recall Z-0993-2013 ยท LZG Pump, Infusion, Insulin

Animas Corporation

Animas(R) 2020 Insulin Infusion Pump Product Usage: This product is indicated for continuous subcutaneous infusion of insulin for the treatment of insulin-requiring diabetes.

DateApr 3, 2013 (initiated Jan 3, 2013)
StatusTerminated
ReasonAnimas has identified a component issue affecting a small supply of the Animas(R) 2020 insulin pumps. The component issue may trigger the pumps to sound a false alarm or warning related to one of the following: - Loss of prime; -Occlusion; or -No cartridge detected.
Root causeNonconforming Material/Component
ActionAnimas sent an Urgent Recall letter dated December 17, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter informed customers that Animas will send a replacement pump free of charge, along with instructions for returning their pump. For questions or concerns contact at 866-796-6373.
Quantity7,157 (494 US pumps affected)
DistributionWorldwide Distribution - USA (nationwide) including the states of: AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and Puerto Rico and the countries of: Australia,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042873ANIMAS MODEL IR1250 INSULIN INFUSION PUMP
Animas Corp., Dec 10, 2004