Recall Z-0870-2013 ยท DXT Injector And Syringe, Angiographic

Medrad Inc

MEDRAD Hand Controller Sheath, Catalog number AVA 500 HCS/3010903, for cardiac catheterization procedures. The Sterile Sheath is specifically intended for use in the x-ray angiography environment to cover the Medrad Avanta Sterile Hand Controller to prevent cross contamination between patients.

DateFeb 21, 2013 (initiated Nov 5, 2012)
StatusTerminated
ReasonThe seal between the white polyethylene and the clear plastic may be missing on the Hand Controller Sheath package, resulting in a potential breach of the sterility of this accessory.
Root causePackaging
ActionMedrad sent a Recall letter dated November 5, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customer were instructed to review their inventory and quarantine any packages with an expiration date prior to April, 2017. Customers were also given the option of conducting a visual inspection on site to determine if the product should be returned. Alternatively, the customer has the option of returning the affected product. If the customer has affected product, they are instructed to contact Medrad Customer Su
Quantity832,175
DistributionWorldwide Distribution including USA nationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050456MEDRAD AVANTA FLUID MANAGEMENT INJECTION SYSTEM
Medrad, Inc., Jun 17, 2005