Recall Z-0841-2013 ยท MUJ System, Planning, Radiation Therapy Treatment

Elekta, Inc.

Monaco RTP System. Radiation treatment planning.

DateFeb 20, 2013 (initiated Feb 11, 2013)
StatusTerminated
ReasonThe beam is displayed at the "gantry" angle, but the dose is calculated at the "rotation_start" angle.
Root causeNonconforming Material/Component
ActionElekta provided customers with an "Important Safety Notice - Dose Display Not Matching Beam Geometry - Monaco RFC 35748". The notice identified the product, problem, and actions to be taken by the customers. Contact the form at 800-878-4267 for questions relating to this notice.
Quantity13
DistributionNationwide Distribution-including the states of AK, CA, FL, NJ, NC, OH, OR, TN, TX, WA, and WI.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110730MONACO RTP SYSTEM