Recall Z-0919-2013 ยท LXH Orthopedic Manual Surgical Instrument

Stryker Howmedica Osteonics Corp.

Command Instrument Broach Handle. Used in Total Hip Arthroplasty. Catalog Number: 6266-0-100

DateMar 8, 2013 (initiated Dec 20, 2012)
StatusTerminated
ReasonReports of handle fracture and subsequent disassociation of cam assembly subcomponents during use.
Root causeProcess control
ActionStryker Orthopaedics sent an "URGENT MEDICAL DEVICE RECALL NOTIFICATION" dated December 20, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at (201) 972-2100 for questions regarding this issue.
Quantity48 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.