Recall Z-0935-2013 ยท LNH System, Nuclear Magnetic Resonance Imaging

GE Healthcare, LLC

Optima MR450w, Discovery MR450, Discovery MR750, Signa HDx, Signa HDxt, Signa HDi, Signa Vibrant, and GE 1.5T Signa HDe MR System. The GE Signa HDx MR system is a whole body magnetic resonance scanner.

DateMar 11, 2013 (initiated May 10, 2011)
StatusTerminated
ReasonAn error on the SR Viewer Reporting Tool in which edits can be made on a report and can be saved into incorrect reports.
Root causeSoftware design
ActionGE Healthcare sent an Urgent Field Safety Notice dated May 5, 2011, to all affected customers. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information. Customers were instructed to open only one patient report at a time when editing a report in SR Viewer. US Customers did not receive a customer letter, however all US systems have been updated. For questions customers were instructed to contact their local Sales/Service Representative. For questions regarding this recall call 262-513-4122.
Quantity4026 (1672 US; 2354 OUS)
DistributionWorldwide Distribution - USA (All States in continental USA including PR except RI.). Internationall to: ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, AZERBAIJAN, BRAZIL, BELARUS, BELGIUM, BULGARIA, BOSNIA & HERZEGOVINA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, EGYPT, ESTONIA, FI
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052293GE 1.5T AND 3.0T SIGNA HDX MR SYSTEM
Ge Medical Systems, LLC, Sep 21, 2005
K052978GE 1.5T SIGNA HDE MR SYSTEM
Ge Medical Systems, LLC, Nov 9, 2005
K081028SIGNA 3.0T MR750 SYSTEM
Ge Medical Systems, LLC, Apr 25, 2008
K083147DISCOVERY MR450
Ge Medical Systems, LLC, Nov 4, 2008
K091536OPTIMA MR450W
Ge Medical Systems, LLC, Jul 17, 2009