Recall Z-1784-2014 ยท LOX Catheters, Transluminal Coronary Angioplasty, Percutaneous

Cordis Corporation

Cordis EMPIRA RX PTCA Dilatation Catheter Catalog # 85R30300S For cardiovascular use.

DateJun 11, 2014 (initiated Feb 5, 2013)
StatusTerminated
ReasonDilatation catheters could exhibit radial versus axial tears should they burst during inflation.
Root causeProcess design
ActionAn Urgent Medical Device Recall letter and Acknowledgement form was sent overnight to mulitple contacts in each account February 7, 2013. A representative will follow-up as necassary to facility obtaining signature, faxing the acknowledgment form to Cordis, collecting and returning units.
Quantity80 units
DistributionWorldwide distribution: US states: AZ, FL, IL, LA, MA, NJ, OR, SC, and TX. Armenia, Austria, Belgium, Colombia, Czech Republic, France, Hungary, Iran, Israel, India, Italy, Kuwait, Latvia, Lebanon, Luxembourg, Mexico, Morocco, Poland, Portugal, Russia, Saudi Arabia, South Africa, Spain, Switzerlan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110133EMPIRA RX PTCA DILATATION CATHETER, EMPRIA NC RX PTCA DILATATION CATHETER
Creganna-Tactx Medical, Jun 10, 2011