Recall Z-0930-2013 ยท IYO System, Imaging, Pulsed Echo, Ultrasonic

Siemens Medical Solutions USA, Inc.

ACUSON S2000 Automated Breast Volume Scanner (ABVS); Model number 10042541 Distributed by Siemens Medical Solutions USA Inc. business Unit: Ultrasound. The S2000 Automated Breast Volume Scanner (ABVS) Ultrasound System for imaging of the breast allows volumetric acquisition of the breast image.

DateMar 9, 2013 (initiated Jan 18, 2013)
StatusTerminated
ReasonOn some ACUSON S2000 ABVS systems, the inside of the transducer pod may contain sharp edges.
Root causeProcess design
ActionSiemens sent a Customer Safety Advisory notification letter to Siemens Service Personnel on January 18, 2013 and which would be provided to all affected sites during their visit to correct the problem. The notification identified the product, the problem and the action to be taken. The local Customer Service Engineer will perform the test at their site. If the system passes the test, no further action is required. If the ABVS pod fails the inspection, the Service Engineer will correct the sharp edge. For questions regarding this recall call 650-694-5398.
Quantity214 affected devices
DistributionWorldwide Distribution - USA including WV and OH; Internationally to Australia, Austria, Bahrain, Belgium, Brazil, China (P.R.) Estonia, France, Germany, Greece, India, Indonesia, Israel, Italy, Japan, Jordan, Kuwait, Lebanon, Macedonia, Malaysia, Mexico, Netherlands, Peru, Philippines, Poland, Port
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081148SIEMENS ACUSON 2000 ULTRASOUND SYSTEM WITH ABVS TRANSDUCER
Siemens Medical Solutions USA, Inc., Ultrasound DI, Dec 22, 2008