Recall Z-1146-2013 ยท KRA Catheter, Continuous Flush

Edwards Lifesciences, LLC

Pressure Monitoring sets with VAMP Plus closed blood sampling system model # T001691A, PXVP0566, PXVP2260, PXVP0826, PXVP0827, PXVP2284AT3, PXVP23X3AT3, T398807D, T100605B, T001762A, T391T00A, VP2. Blood sampling system used on patients requiring periodic withdrawal of blood samples.

DateApr 18, 2013 (initiated Jan 30, 2013)
StatusTerminated
ReasonThe device is being recalled because the firm has received multiple complaints related to incorrect assembly that may result in dilution of blood samples.
Root causeProcess change control
ActionThe firm initiated a recall in Europe via customer letters starting January 30, 2013 to notify the recall of model number T001691A lot number 59326333. The recall was expanded and a worldwide recall via customer letters was initiated on February 11, 2013. The letter identified the product, problem, and actions to be taken by the customers.
Quantity30,467
DistributionWorldwide Distribution-USA (nationwide) including the states of AR, CA, FL, HI, ID, IN, KS, KY, LA, MA, MI, MN, MO, NC, ND, NM, NV, NY, OH, OK, PA, SD, TN, TX, VA, WA, and WI) and Europe.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K896819VAMP VENOUS/ARTERIAL BLOOD MANAGEMENT PROTECTION
Baxter Healthcare Corp, Jan 19, 1990
K925638PHOENIX DISPOSABLE PRESSURE TRANSDUCER #73-600
Baxter Healthcare Corp, Oct 18, 1993