ClearanceBrief
Recall Z-0773-2013 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Zimmer, Inc.

Trilogy Acetabular Shell with Cluster Holes. The Trilogy Acetabular System is indicated for either cemented or uncemented use in individuals undergoing surgery for rehabilitating hip damage.

DateFeb 4, 2013 (initiated Sep 14, 2012)
StatusTerminated
ReasonSome units of Trilogy shells manufactured since March 2009 may be below the lower range of the pore size specification (as low as 28% vs. 30% porosity specification). On 10/31/2012, the firm expanded the recall action to include lots of Uni-Hole Trilogy Acetabular System shells.
Root causeProcess control
ActionZimmer sent an "URGENT MEDICAL DEVICE NOTIFICATION" letter dated September 20, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers.The firm expanded the scope of the recall in January 2013 to include devices that use a secondary mean of fixation. New recall letters will be submitted to distributors and surgeons.
Quantity275,952 units
DistributionWorldwide Distribution-USA (nationwide) and the countries of Denmark, Germany, Greece, Hungary, Ireland, Italy, Netherlands, Spain, Sweden, UK, Poland, France, Australia, Chile, Canada, India, Japan, New Zealand, Peru, South Africa, Colombia, Ukraine, and UAE.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K934765TRILOGY ACETABULAR SYSTEM
Zimmer, Inc., Apr 29, 1994