Recall Z-0868-2013 ยท DQY Catheter, Percutaneous

Vascular Solutions, Inc.

Vascular Solutions, SuperCross Microcatheter, Sterilized using ethylene oxide, Rx Only, Model numbers 5300 - 5307, 5324, 5325, 5329, 5332, 5336, 5340, and 5341. Product Usage: The SuperCross catheter is intended to be used in conjunction with steerable guidewires to access discrete regions of the coronary and/or peripheral vasculature. It may be used to facilitate placement and exchange of guidew

DateFeb 20, 2013 (initiated Jan 29, 2013)
StatusTerminated
ReasonVascular Solutions, Inc has discovered a potential problem with a number of lots of Supercross microcatheters. A potential for surface contamination on aged catheters was observed during in-house testing.
Root causeMaterial/Component Contamination
ActionVascular Solutions sent an "Urgent Medical device Recall" letter dated January 28, 29013 to all affected customers. The letter identified the affected product, described the problem and actions to be taken. The letter advised customers to remove the product from their inventory and to complete and return the VSI Account Inventory Form. For questions contact your local Sales Representative.
Quantity4731
DistributionWorldwide Distribution - USA Nationwide including the states of AL, AZ, AR, CA, CO, CT, DE, CD, FL, GA, HI, ID, IL ,IN, IA, KY, LA, ME, MD, MA, MI,MN, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, SC, SD, TN, TX, UT, VT VA, WA, WV, WI, WY and AUSTRIA, CANDA, DENMARK, FINLAND, FRANCE, GERMA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101659SUPERCROSS MICROCATHETER (130 CM), SUPERCROSS MICROCATHETER (150 CM) MODEL 5300, 5301
Vascular Solutions, Inc., Dec 16, 2010