Recall Z-0956-2013 ยท HRS Plate, Fixation, Bone

Synthes USA HQ, Inc.

Variable Angle Locking Compression Plate/VA-LCP Dorsal Distal Radius Plate, Part Number 02.115.431. Indicated for: Dorsally displaced fractures, Open joint reconstruction, and Corrective osteotomies after distal radius malunion.

DateMar 15, 2013 (initiated Jan 11, 2013)
StatusTerminated
ReasonIt was discovered that one of the VA-LCP Dorsal Distal Radius Plates was incorrectly etched as Part Number 02.115.431 (110 degree) instead of the correct Part Number 02.115.231 (90 degree).
Root causeLabeling mix-ups
ActionThe firm initiated their recall of this product on January 11, 2013 by sending an Urgent Notice: Medical Device Recall letter to consignees. The letter identified the affected product and the issue. Synthes requested customers to examine their inventory for the affected product, remove them from use, and return them to Synthes. Additionally, customers were to follow the steps provided in the event that they did or did not have any affected product on hand. Questions should be directed to 610-719-5450 or FieldAction@synthes.com.
Quantity4
DistributionDistributed in the states of IN, OH, PA, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K1026942.4MM VA-LCP DORSAL DISTAL RADIUS PLATES
Synthes (Usa), Dec 9, 2010