ClearanceBrief
Recall Z-0791-2013 ยท MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate

Medacta Usa Inc

Medacta Quadra Trial Broach size 0 part # 01.10.10.045, for use with the Medacta Total Hip Prosthesis System. Product Usage - used to broach the intramedullary canal of the femur in order to create space for the femoral stem during a hip replacement surgery.

DateFeb 7, 2013 (initiated Dec 17, 2012)
StatusTerminated
ReasonThe tip of the Quadra Trial Broach size 0, broke during a surgery.
Root causeDevice Design
ActionMedacta sent a Urgent Safety Notification letter dated December 13, 2012 to all affected customers. The letter identified the recalled product, problem and actions to be taken. Customers were instructed to check inventory for affected product and contact QA@medacta.us.com for an RMA# to return items to Medacta USA. A new version of the recalled product will be provided to all affected customers. For questions call 805-437-7085 x26.
Quantity26 pieces U.S.A. (226 pieces Worldwide)
DistributionUS Nationwide Distribution including the states of Arizona, California, Idaho, Illinois, North Carolina, Ohio, Utah, Washington and US Virgin Islands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K072857TOTAL HIP PROSTHESIS SYSTEM QUADRA S + COCRMO FEMORAL BALL HEAD + APRICOT
Medacta International S.A., Feb 4, 2008
K082792MEDACTA TOTAL HIP PROSTHESIS SYSTEM - QUADRA H AND R FEMORAL STEMS
Medacta International S.A., Dec 29, 2008
K083558MEDACTA TOTAL HIP PROSTHESIS SYSTEM - QUADRA C
Medacta International S.A., Mar 6, 2009
K093944MEDACTA TOTAL HIP PROTHESIS SYSTEM-AMISTEN H , QUADRA H, QUADRA S
Medacta International S.A., Jan 21, 2010
K103189MEDACTA TOTAL HIP PROSTHESIS SYSTEM - AMISTEM C & QUADRA C SHORT NECK
Medacta International S.A., Dec 20, 2010