Recall Z-2164-2013 ยท FRN Pump, Infusion
B Braun Medical, Inc.
Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primary and secondary fluids when positive pressure is required. The infusion system is capable of delivering fluid from a negative head height (when the IV fluid container is lower than the pump), and provides clinically accepted volumetric accuracy for all standard IV fluids, includi
| Date | Sep 6, 2013 (initiated Oct 31, 2012) |
| Status | Terminated |
| Reason | This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on 26 August, 2011 due to an issue in the Outlook ES Safety Infusion System where the Care Area can inadvertently be exited thereby removing DoseGuard dose limit protection. This occurs when a sequence of HOLD - HOLD key strokes are performed on the pump key panel while the user is in the Dose Dat |
| Root cause | Software design |
| Action | The firm decided to conduct a field correction to correct the issue and carried it out through routine field services. The firm was notified by FDA that this is a reportable event, so they notified their consignees via letters dated 10/31/2012. |
| Quantity | 1,970 pumps added for this expansion |
| Distribution | Nationwide Distribution. |
| FDA record | Open on fda.gov |