ClearanceBrief
Recall Z-1450-2013 ยท KPR System, X-Ray, Stationary

Philips Healthcare Inc.

Philips BuckyDiagnost Radiographic. Catalog Numbers: 704031, 704032, 704035, 704060, 704062 This system is used for making x-ray exposures for diagnostics

DateJun 17, 2013 (initiated Jan 16, 2013)
StatusTerminated
ReasonAn inspection of IATD (Installation Acceptance Test Document)IA SID (Source Image Distance) records shows table height measurements out of specification with the consequence of potential system mis-alignment
Root causeRelease of Material/Component prior to receiving test results
ActionPhilips Healthcare issued an Urgent Field Safety Notification on January 11, 2013, to advised users of issued and perform corrective action to the units. Field Service Engineer (FSE) will execute corrections and attach record to IATD file. Your proposed corrective action plan (CAP) for the Bucky Diagnost system involves two mandatory field actions including new testing for sites with missing documentation and correcting the table height measurements for systems where the incorrect measurement procedure was used. You have also revised your IATD document to improve instructions to the field en
Quantity116 units
DistributionWorldwide Distribution - US Nationwide and the country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K945278PHILIPS BUCKY DIAGNOST FAMILY
Philips Medical Systems North America, Inc., Dec 14, 1994