Recall Z-0977-2013 ยท CFR Hexokinase, Glucose

Mindray DS USA, Inc. d.b.a. Mindray North America

BS-200 Chemistry Analyzer Designed for clinical laboratory use.

DateMar 22, 2013 (initiated Nov 28, 2012)
StatusTerminated
ReasonAn issue has been identified with the BS-200 where clinical chemistry results may be inaccurate.
Root causeDevice Design
ActionMindray North America /Mindray DS USA Inc. sent Urgent Product Corrective Action letters via certified mail, return receipt requested, signature required on 11/28/2012. The letters identified the affected product as well as the issue, effect on patients, and the corrective action. Customers are to remove the analyzer from service until the affected pump has been replaced. Also, customers are to contact Mindray Customer Service Representative Tony Li at 1-800-288-2121 x2112 for technical assistance, and to arrange for replacement pumps at no charge.
Quantity3 units
DistributionDistributed in the states of NC and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022519LDL CHOLESTEROL (AUTOMATED)
Jas Diagnostics, Inc., Sep 30, 2002
K072018BS-200 CHEMISTRY ANALYZER, MODE BS-200
Shenzhen Mindray Bio-Medical Electronics Co., Ltd., Feb 1, 2008