Recall Z-0745-2013 ยท JIS Calibrator, Primary

Siemens Healthcare Diagnostics, Inc.

Dimension(R) IRON Calibrator (DC85), Siemens material number 10445010. The Dimension(R) IRON Calibrator is an in vitro diagnostic product intended to be used to calibrate the Iron method for the Dimension(R) clinical chemistry system.

DateJan 29, 2013 (initiated Dec 20, 2012)
StatusTerminated
ReasonSiemens Healthcare Diagnostics has received complaints of IRON calibrations outside of acceptable calibration slope (m) values (0.97 - 1.03) when using IRON calibrator lots 2DN060 and 2FN063. Lots 2GN068 and 2JN074 may also be impacted but no complaints have been received.
Root causeNonconforming Material/Component
ActionSiemens sent a "CUSTOMER NOTIFICATION" letter dated December 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact Siemens Technical Solutions Center at 1-800-441-9250 or your local Siemens Healthcare Diagnostics representative for assistance with this issue.
Quantity5999
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060266DIMENSION IRON CALIBRATOR
Dade Behring, Inc., Mar 9, 2006