Recall Z-0901-2013 ยท JAK System, X-Ray, Tomography, Computed

GE Healthcare, LLC

GE Healthcare Volume Viewer Plus 5.x (Advantage Workstation 4.2) Volume Viewer 3.x (Advantage Workstation 4.1) Volume Analysis (Advantage Workstation 4.0) Volume Analysis (Advantage Workstation 3.1) Advantage Windows Volume Rendering Option is intended to provide fast, easy, volume visualization of three dimensional structures imaged with a computed tomography (CT) or magnetic resonance (MR

DateMar 1, 2013 (initiated May 25, 2012)
StatusTerminated
ReasonIncorrect views and measurements possible. GE Healthcare became aware of multiple potential safety issues involving the Advantage workstation, Volume Viewer Plus and Volume Analysis.
Root causeSoftware design
ActionGE Healthcare sent a "Urgent Medical Device Correction " letter dated May 23, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Customers were instructed to discontinue use of the Edit Contour feature until this software functionality is corrected or replaced. GE Healthcare will correct all affected systems by providing a software upgrade at no cost to you. A GE Healthcare service representative will contact you to arrange for this correction. If you have any questions or concerns regarding this notification, p
Quantity363
DistributionWorldwide Distribution - USA including the states of AR, CA, CO, FL, GA, IL IN, IA, LA, MD, MA, MI, MN, MO, NJ, NM, NY, NC, OH, OK, PA, PR, SC, SD, TN, TX and VA. and the countries of ALGERIA, AUSTRALIA, AUSTRIA, ARGENTINA, BELARUS, CANADA, CHINA, CHILE, EGYPT CZECH REPUBLIC, FRANCE, FINLAND, GERMA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041521VOLUME VIEWER PLUS
General Electric Medical Systems, Jun 22, 2004
K972399ADVANTAGE WINDOWS VOLUME RENDERING OPTION
Ge Medical Systems, Inc., Sep 23, 1997