ClearanceBrief
Recall Z-0670-2013 ยท DWF Catheter, Cannula And Tubing, Vascular, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Sarns high-flow aortic arch cannula, 6.5 mm (20 Fr) OD with 3/8in connector, 11in (28 cm) long The Sarns" High Flow Aortic Arch Cannula and Sarns" Flexible Arterial Cannulae are indicated for use in perfusion of the ascending aorta during cardiopulmonary bypass surgery

DateJan 11, 2013 (initiated Jan 7, 2013)
StatusTerminated
ReasonFlexible Arterial Cannula and High-Flow Aortic Arch Cannula connectors were identified to contain flash on the exterior of the cannula connector that exceeded established specifications. Because the flash has the potential to detach in the operative field, TCVS has decided to remove all affected lots that were manufactured with the affected connector part.
Root causeProcess control
ActionTERUMO sent an URGENT MEDICAL DEVICE RECALL notification to all consignees on January 7, 2013. Terumo CVS is advising users to discontinue use and return all affected cannulae in inventory. Terumo CVS will issue credit for returned cannulae. Terumo CVS providied a list of marketed devices that can be used as alternatives. Contact Terumo CVS Customer Service with questions or concerns: 800- 521- 2818 Monday Friday, 8 AM - 6 PM EST For questions regarding this recall call 734-741-6173.
Quantity6830 total all products
DistributionNationwide and worldwide: AL , AR , AZ , CA , CO , CT , FL , GA , HI , IA , IL , IN , KS , KY , LA , MA , ME , MI, MO , NC , ND , NV , NY , OH , OK , PA , SD , TN , TX , UT , VA , WA , and WI. Bahrain, BRAZIL, CANADA, CHILE, Honduras, Singapore, Taiwan, Trinidad and Tobago, and UNITED ARAB EMIRATES
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K083301X-COATED TERUMO AND SARNS DUAL STAGE VENOUS RETURN CATHETERS, SARNS MALLEABLE VENOUS AND DUAL STAGE RETURN CATHETERS
Terumo Cardiovascular Systems Corp., Dec 18, 2008
K770429CANNULAE, AORTIC, ARCH, HIGH FLOW
3M Health Care, Sarns, Mar 16, 1977
K771499FLEXIBLE ARTERIAL CANNULAE
3M Health Care, Sarns, Aug 16, 1977