Recall Z-0785-2013 ยท IOR Wheelchair, Mechanical

Invacare Corporation

Solara 3G Custom Manual Wheelchair Product Usage: To provide mobility to a person restricted to a sitting position.

DateFeb 4, 2013 (initiated Sep 10, 2012)
StatusTerminated
ReasonThe potential exists for the wheel to rotate freely despite engagement of hub brake.
Root causeProcess control
ActionInvacare sent an URGENT: RECALL INFORMATION Letter dated September 10, 2012 to all affected customers. The letter identified the affected product, problem and actions to be taken. The letter instructed customers to 1) Complete the Provider Response Card indicating your receipt and understanding of these instructions. 2) Contact your customers to inform them of the recall. It is recommended that you provide them with the enclosed Customer Letter. Set up an appointment to correct the customer's wheelchair as soon as possible. 3) Please examine the wheelchair to confirm it has one of the se
Quantity3 units
DistributionUS Nationwide Distribution including the states of CA & MT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K984447MODEL ACTION AT II MANUAL WHEELCHAIR
Invacare Corp., Jan 8, 1999