Recall Z-0362-2017 ยท FRN Pump, Infusion

Hospira Inc.

Symbiq Two Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids. List number 16027. Note that Hospira retired this pump June 30, 2015. The Hospira Symbiq Infusion System is intended for the delivery of fluids, solutions, drugs, agents, nutritionals, electrolytes, blood, and blood products via parenteral, enteral, intravenous, intra-arterial, subcutaneous, epidural,

DateMar 16, 2011
StatusTerminated
ReasonMay go to a white screen due to a problem with the SOM2 module. A white screen is a diagnostic screen displayed to the user as part of the safety shutdown procedure of the device. The infusion and the processor stop, sound a 3 minute audible alarm from the backup buzzer and deactivate the couch screen and the automated loading/unloading of cassettes. Delay or interruption of therapy.
Root causeDevice Design
ActionHospira sent an Urgent Device Recall Letters, dated February 5, 2013, to all affected customers. Customers were directed to notify their consignees if they further distributed the product and confirm they have done so by return of the reply form. Healthcare professionals were to contact Stericycle. Hospira discontinued support of these devices as of June 30, 2015. Customers with questions were instructed to call 1-800-201-9064. For questions regarding this recall call 224-212-4892.
Quantity10897
DistributionWorldwide Distribution - US (nationwide) and Canada