ClearanceBrief
Recall Z-0721-2013 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

ConMed Corporation

Goldvac(TM) UltraClean(R) Accessory Electrode 6 inch Coated Blade with Extended Insulation, REF/Catalog No. 60-7523-001, Sterile, Rx ONLY -- CONMED CORPORATION, Electrosurgical, Cutting and Coagulation and Accessories.

DateJan 24, 2013 (initiated Dec 27, 2012)
StatusTerminated
ReasonConMed received complaints of some devices which had broken through the seal of the sterile pouch. ConMed confirmed instances where the pouch seal was compromised on the affected products. In no instance has it been reported to ConMed that a compromise in the sterile barrier has resulted in illness or injury. The compromised seals were discovered prior to use.
Root causeEquipment maintenance
ActionConMed sent an Urgent Field Safety Notice dated January 7, 2013, and Business Reply Form to all affected customers via USPS Priority Mail. International notices were sent via FedEx International Priority on December 27, 2012. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to review their inventory and complete Attachment II and return it with unused affected product to: ConMed Corporation 525 French Road, Utica, NY 13502 Customer were asked to return Attachment II even if they don't have any affected product. For quest
Quantity313,190 (total for all devices)
DistributionWorldwide Distribution - USA (nationwide) Puerto Rico and internationally to Belgium, Colombia, Czech Republic, Dubai, Guatemala, Hong Kong, India, Israel, Italy, Japan, Kenya, Kuwait, Malaysia, Mexico, New Zealand, Norway, Oman, Pakistan, Portugal, Saudi Arabia, Thailand, and Uruguay.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K991855CONMED STEALTH NON-STICK ACCESSORY ELECTRODE, MODEL 138XXX
Conmedcorp, Jun 15, 1999