Recall Z-0700-2013 ยท GEX Powered Laser Surgical Instrument

AMS Innovative Center - San Jose

GreenLight HPS (High Performance System) GreenLight XPS (Xcelerated Performance System) Manufactured by AMS, Solutions for Life Innovation Center. GreenLight HPS and XPS Surgical Lasers Systems and accessories are intended for use in endoscopic 532 nm laser resection of the prostate for the treatment of benign prostatic hyperplasia.

DateJan 22, 2013 (initiated Dec 11, 2012)
StatusTerminated
ReasonDuring electrical recertification testing it was found that the Voltage Scaling Board was non-conforming to the IEC 60601-1-2 standard.
Root causeNonconforming Material/Component
ActionAMS sent "NOTICE OF DEVICE FIELD CORRECTIVE ACTION" letters dated December 20, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. Contact the firm at 1-408-456-3320 for questions about this notification.
Quantity389 devices, all model numbers.
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062719GREENLIGHT HPS SERIES SURGICAL LASER SYSTEM & ACCESSORIES
Laserscope, Dec 1, 2006
K092735GREENLIGHT XPS LASER SYSTEM
American Medical Systems, Inc., Nov 9, 2009