ClearanceBrief
Recall Z-0877-2013 ยท KMI Monitor, Bed Patient

Intel-GE Care Innovations LLC

QuietCare-Networked product, Facility Server/model: QC101601; The QuietCare system is intended to detect and analyze residents' movements.

DateFeb 23, 2013 (initiated Dec 16, 2012)
StatusTerminated
ReasonCare Innovations detected a software anomaly: During the engineering investigation into the cause of why a Wander Alert was not generated, a defect in the implementation of the Wander and Night Motion ADLs was discovered.
Root causeSoftware design
ActionCare Innovations sent a "DEVICE CORRECTION NOTIFICATION" letter dated 12/20/12 to all affected customers. The letter identified the product and problem. Correction was performed remotely by the firm. Contact the firm at 917-232-4775 for questions regarding this notice.
Quantity19 units
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.