Recall Z-0688-2013 ยท JLA Conversion To Creatinine, Creatine

Perkinelmer

NEONATAL TOTAL GALACTOSE KIT Product Numbers: 3029-0010, 3029-001B. This kit is intended for the quantitative determination of total galactose (galactose and galactose-1-phosphate) concentrations in blood specimens dried on filter paper as an aid in screening newborns for galactosemia.

DateJan 16, 2013 (initiated Nov 26, 2012)
StatusTerminated
ReasonTruncated V-Bottomed Microplate Wells used in assays contain defective wells.
Root causeProcess control
ActionPerkin Elmer initiated the recall on November 23, 2012. The Recall Letter will be delivered to the end customer by email facilitated through our PerkinElmer Technical Support group located in Akron Ohio. Technical Support will be asked to make sure the customers receive the Recall Letter and implement the recommended precautions: - Inspect the wells of row A and row H of the affected microplate lots for possible holes and cracks prior to use. - If a hole or crack is visually detected in a well of the microplate, the well should not be used for testing. - If a hole or crack is not visually d
Quantity3029-0010 =1562 kits (10 plats/kit) and 3029-001B=13 kits (50plates/kit)
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071649NEONATAL TOTAL GALACTOSE KIT; NEONTAL TOTAL GALACTOSE KIT, MODELS 3029-0010 AND 3029-110B
Wallac OY, Aug 7, 2008