Recall Z-0686-2013 ยท JNB Ninhydrin And L-Leucyl-L-Alanine (Fluorimetric), Phenylalanine

Perkinelmer

NEOGRAM AMINO ACIDS AND ACYLCARNITINES TANDEM MASS SPECTROMETRY KIT Product Number: 3026-0030. This kit is intended for the measurement and evaluation of amino acid, free carnitine, and acylcarnitine concentrations from newborn heel prick blood samples dried on filter paper. (Table 1 of Attachment 2) details the analytes measured by the kit. Quantitative analysis of amino acids, free carnitine a

DateJan 16, 2013 (initiated Nov 26, 2012)
StatusTerminated
ReasonTruncated V-Bottomed Microplate Wells used in assays contain defective wells.
Root causeProcess control
ActionPerkin Elmer initiated the recall on November 23, 2012. The Recall Letter will be delivered to the end customer by email facilitated through our PerkinElmer Technical Support group located in Akron Ohio. Technical Support will be asked to make sure the customers receive the Recall Letter and implement the recommended precautions: - Inspect the wells of row A and row H of the affected microplate lots for possible holes and cracks prior to use. - If a hole or crack is visually detected in a well of the microplate, the well should not be used for testing. - If a hole or crack is not visually d
Quantity180 kits (20 plates/kit)
DistributionNationwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021541NEOGRAM PKU TANDEM MASS SPECTROMETRY KIT
Perkinelmer Life Sciences, Oct 28, 2002