Recall Z-0747-2013 ยท JJY Multi-Analyte Controls, All Kinds (Assayed)

Alere San Diego, Inc.

Alere Cholestech LDX Calibration Verification, 4 vial set, Catalog Number: 88770 Alere Cholestech LDX Calibration Verification (hereafter referred to as CAL VERS) enables users to monitor the performance of total cholesterol (TC), high density lipoprotein cholesterol (HDL), triglycerides (TRG), and glucose (GLU) test procedures on the Alere Cholestech LDX System. Calibration verification is optio

DateJan 29, 2013 (initiated Oct 22, 2012)
StatusTerminated
ReasonOut of range (increased) HDL Cholesterol results.
Root causeComponent design/selection
ActionA recall letter was sent on 10/22/12 to all customers who purchased the Alere Cholestech LDX Mutianalyte Control, 2 vial set, Catalog Number 88769, Lot #C2510 & Lot #C2510A. The letter informed the customers that these lots of controls may yield results which are out of range high for the Level 2 control when tested with the Alere Cholestech LDX HDL test. Customers are instructed of the actions to be taken. On 11/16/12, the recall expanded to include the Alere Cholestech LDX Calibration Verification, 4 vial set, Catalog 88770, Lot#271097. The problem identified is that these lots of calibr
Quantity614 kits
DistributionWorldwide Distribution - USA (nationwide) Puerto Rico and South Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K102700LIPID CONTROLS; MULTI- ANALYTE CONTROL; MULTI-ANALYTE CALIBRATION VERIFICATION MATERIALS
Alere, San Diego, Dba Biosite INcorporated, Dba IN, Dec 21, 2010