Recall Z-0660-2013 ยท KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented

Biomet, Inc.

PT113950 PT HYBRID Glenoid Post Regenerex porous titanium construct Sterile R Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented

DateJan 9, 2013 (initiated Dec 6, 12)
StatusTerminated
ReasonBiomet is recalling Part Number PT-113950 PT Hybrid Glenoid Post, following an investigation which identified that the male thread of the post may be oversized. This oversized condition can vary in degree and may lead to the following three events: 1) If the PT Hybrid Glenoid Post is not fully seated into the Hybrid Base and the implant construct is implanted, then a gap of 1-3 mm will be prese
Root causeNonconforming Material/Component
ActionBiomet Inc. sent a Urgent Medical Device Recall Notices dated December 6, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The notice identified the reason for the recall including directions to immediately locate, discontinue use of the device. Surgeons were asked to continue monitoring of patients implanted. Customers were asked to confirm receipt of the notice by calling 1-888-348-9500 extension 1570. Customers with product to be returned are asked to obtain RGA # by calling 574-372-6677 . A FAX Back Re
Quantity1050
DistributionWorldwide Distribution -- US (nationwide) including the countries of , Canada, Japan, Argentina, Australia ,Netherlands, Europe and Chile.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060694MODULAR HYBRID GLENOID
Biomet Manufacturing Corp, Dec 11, 2006